How is the reagent blank entered when using the open chemistry option with a 961X Silica/Phosphate analyzer?

ID documento

ID documento TE12617

Data pubblicazione

Data pubblicazione 15/02/2022
Domande
How is the reagent blank entered when using the open chemistry option with a 961X Silica/Phosphate analyzer?
Sommario
961X Silica/Phosphate analyzer reagent blank
Risposta
When using Hach prepared reagents on any Polymetron 9610 Open Chemistry Analyzer or manual entry of Reagent Blank (same as Zero entry) with User prepared reagents, follow the steps below:  
1. Press the CAL key to display the CALIBRATE MENU.
2. Select “OVERRIDE CAL” then press the ENTER key.  
3. From the CANCEL AUTO CAL? menu select “✔YES” at the bottom of the screen and then press the ENTER key.
4. From the OVERRIDE CALIBRATION menu select “ZERO” and then press the ENTER key.
5. Enter the desired Zero value…..
 
                                    + 0 0 1 . 0 0 0   Example
 
6. Press the ENTER key and then the HOME key.
 