Domande
How to configure a Warning relay on the SC200 for CL17sc 'low sample flow'?
Sommario
sc200 relay for CL17sc low sample flow
Risposta
The “LOW SAMPLE FLOW” warning from the CL17sc can be associated with a relay to provide feedback on loss of flow.
This is performed by selecting: MENU-->SC200 SETUP-->RELAY SETUP-->(choose the right relay A, B, C, or D)-->SET FUNCTION-->WARNING and press enter.
Navigate back to RELAY SETUP MENU -->ACTIVATION --> Check mark (LOW SAMPLE FLOW)
This is performed by selecting: MENU-->SC200 SETUP-->RELAY SETUP-->(choose the right relay A, B, C, or D)-->SET FUNCTION-->WARNING and press enter.
Navigate back to RELAY SETUP MENU -->ACTIVATION --> Check mark (LOW SAMPLE FLOW)