What type of probes (digital or analog) can I use with my sc200 controller?
Sommario
Digital and analog sensors attached to the sc200 controller.
Risposta
There are many different types of sc200s that can be purchased from Hach. You have the option of one digital input, two digital inputs, one analog input, two analog inputs, or one analog input and one digital input. The digital inputs cannot be added after the fact, so if the sc200 is purchased with one digital input you will never be able to attach two digital sensors.
You are able to add analog modules (up to two total) at any time. If you have a dual digital sc200 and you add an analog module and sensor, you will still be limited to two sensors and the controller will prompt you to delete the excessive sensor(s).
For a sc200 part number configurator please see the following link: https://www.hach.com/controllers-digital/sc200-controller/family?productCategoryId=35546907050
You are able to add analog modules (up to two total) at any time. If you have a dual digital sc200 and you add an analog module and sensor, you will still be limited to two sensors and the controller will prompt you to delete the excessive sensor(s).
For a sc200 part number configurator please see the following link: https://www.hach.com/controllers-digital/sc200-controller/family?productCategoryId=35546907050