What is the proper procedure for resetting/adjusting the totalizer on the sc200 controller?

ID documento

ID documento TE7577

Data pubblicazione

Data pubblicazione 20/12/2021
Domande
What is the proper procedure for resetting/adjusting the totalizer on the sc200 controller?
Sommario
Procedure for resetting/adjusting when 'Totalizer is Full +++++++++' displaying on screen
Risposta

If the totalizer is set in Auto mode, once the totalizer is full, it will automatically reset. If the volume multiplier is set to Auto it will change from x1, x10, x100 and so on until the max value is reached and then it will Auto reset.

If the totalizer is set in Manual mode, it will require user to select the Reset Totalizer option and manually reset the totalizer.