What speed should the Lachat RP-150 pump be set at when running Lachat methods?
Sommario
Lachat pump speed setting
Risposta
The Lachat RP-150 pump is typically set at a speed of 35 when running methods. Lachat ultra low flow methods run at a pump speed of 10. The ultra low flow methods are easy to recognize as they will start with the number 80.