Should the IntelliCAL™ ISENH3181 Ammonia Ion Selective Electrode (ISE) be stored in Ammonia Storage Solution or 1000 mg/L NH3-N Standard?

ID documento

ID documento TE2678

Data pubblicazione

Data pubblicazione 20/12/2021
Domande
Should the IntelliCAL™ ISENH3181 Ammonia Ion Selective Electrode (ISE) be stored in Ammonia Storage Solution or 1000 mg/L NH3-N Standard?
Sommario
Storage of IntelliCAL™ Ammonia ISE probe
Risposta
For short-term storage, it is currently recommended to store the Ammonia ISE in 1000 mg/L NH 3 -N Standard, without Ionic Strength Adjustor (ISA), instead of the Ammonia Storage Solution. Measurement error and drift is lessened when stored in the 1000 mg/L standard.

For long term storage (over one week of disuse), remove the membrane (this can be allowed to dry), rinse the electrode, and store the electrode in pH Electrode Storage Solution (Catalog Number 2756549).