How often should a calibration check be done on the DR300?

ID documento

ID documento TE6258

Data pubblicazione

Data pubblicazione 06/01/2022
Domande
How often should a calibration check be done on the DR300?
Sommario
Pocket Colorimeter calibration check
Risposta
It is recommended that the calibration on the DR300 should be checked every 3 months using SpecCheck Secondary Gel Standards. Regulatory agencies may have different requirements. 
A primary standard solution should also be used periodically to validate the test procedure, reagents, and instrument. Instructions for performing an Accuracy Check - Standard Solution Method can be found in the Hach method.

This is also true for the Pocket Colorimeter II.