Why isn't the Prognosys information showing up on the 5500 sc analyzer display?

ID documento

ID documento TE5790

Data pubblicazione

Data pubblicazione 27/08/2018
Domande
Why isn't the Prognosys information showing up on the 5500 sc analyzer display?
Sommario
5500 sc and Prognosys information on display.
Risposta
If the analyzer is configured for multi-stream, it will not show up.  The Prognosys data takes up a certain amount of space on the display that is not available if multiple channels are displayed.  If the display is changed to only show a single channel, the Prognosys data will appear.
Push right or left arrow button while on the home screen to scroll through the different display options.