What are the menu steps to clear the "Replace Sensor" Warning on a pH sensor connected to the sc200 analog module?

ID documento

ID documento TE5280

Data pubblicazione

Data pubblicazione 05/01/2022
Domande
What are the menu steps to clear the "Replace Sensor" Warning on a pH sensor connected to the sc200 analog module?
Sommario
Clear warning for pH sensor connected to sc200 analog module.
Risposta
Please select the following menu choices:
Menu> Sensor Setup> pH Diff> Diag/Test> Reset Sensor> (you will see message – Warning Resetting Sensor Days Resets Calibration Data to Defaults) press Enter> Yes (to reset sensor). This will clear the error completely.

See the below articles for more information
Why am I unable to clear Calibration Required or Replace Sensor warnings for pHD or ORP sensor?