How can the chlorine sensor in the CL10 panel be cleaned?

ID documento

ID documento TE4794

Data pubblicazione

Data pubblicazione 06/03/2019
Domande
How can the chlorine sensor in the CL10 panel be cleaned?
Sommario
Cleaning the CL10 chlorine sensor.
Risposta
If you find you need to clean the chlorine sensor (this will be indicated by poor performance) we recommend that you rinse the sensor with deionized water. The cleaning needs to be gentle as to not damage the membrane.

In the instance of Iron or Manganese staining you can clean the sensor and flow cell with a 1:3 mixture of 3-5% hydrogen peroxide solution and vinegar.

If the sample running through the CL10 panel fouls the sensor quickly, it is recommended that a strainer is installed before the sample inlet and an acidification unit (Catalog Number/Order Code LZY051) is used.