What are the specifications for the Power Supply, 115 Vac, 60 Hz, for Portable Ultraviolet Lamps 26708 & 26710?

ID documento

ID documento TE3678

Data pubblicazione

Data pubblicazione 18/09/2018
Domande
What are the specifications for the Power Supply, 115 Vac, 60 Hz, for Portable Ultraviolet Lamps 26708 & 26710?
Sommario
User information and specifications for the Power Supply, 115 Vac, 60 Hz, for Portable Ultraviolet Lamps 26708 & 26710 (Product # 2670700).
Risposta
The following information is from the manufacture UVP® for thier model number PS-11.

INTRODUCTION
The Pen-Ray Power Supplies are made specifically for the Pen-Ray Lamps, these power supplies provide proper voltage and current for starting and operating the lamps. Mating recommendations are available to provide optimum energy output from and maximum life span of the Pen-Ray lamps.

DESIGN
Each power supply is designed to be maintenance free. There is an internal circuit protection fuse which is designed to prevent mains circuit shut down in the unlikely case of a catastrophic internal failure. The connector must be used with the mating connector supplied with the Pen-Ray lamps. The connector is certified to UL and CE compliance for use with this power supply. The power supply carries the CE, ETL and ETLc marks.

OPERATING INSTRUCTIONS
The operation of the PS series power supply is very simple:
• Make sure the switch is in the off position.
• Connect the Pen-Ray lamp to the connector making sure the connector is completely inserted into its mating part.
• Plug the unit into proper mains power supply.
• Turn on the unit and the lamp should immediately light up.
• Always unplug the unit when disconnecting or connecting the lamp.

MAINTENANCE
Keep the connector clean and free from debris, oil, fluids, etc. If cord should become frayed or cracked they should be replaced immediately. If connector arcing should occur due to dirt in the connector or attempting to operate a Pen-Ray lamp which operates in excess of 300volts. Do not continue operation. The unit should be returned to the factory for repair. (Cleaning the carbonized area between the connector pins may restore function to the power supply).

WARNINGS
• Never handle the connector when the unit is in the position as hazardous shock condition will exist.
• Do not attempt to operate lamps with operating voltages greater than 300 volts continuously. Starting voltages do not apply.
• If the equipment is used in a manner not specified by UVP, Inc. the protection provided by the equipment may be impaired.

Unit may become HOT during continuous use.

If internal circuit protection fuse blows, the unit should be returned to the factory for service. The internal protection fuse is designed to trip if the primary coil short circuits for any reason protecting the user and the circuit it is plugged into. The unit may be cleaned with any mild cleaning agent. Avoid acids and caustic substances although powder coated finish is extremely durable. Do not use solvent cleaners on plastic end caps.

ENVIRONMENTAL CONDITIONS
• For indoor use only
• Temperature 5 to 40OC
• Maximum on relative humidity 80% for temperatures up to 31OC decreasing linearly to 50% relative humidity at 40OC
• Mains supply voltage fluctuations not to exceed +10% of the nominal voltage
• Installation is category II
• Pollution degree is 2

SPECIFICATIONS
Primary Voltage: 115 VAC/60Hz
Primary Current: 0.40 A
Starting Voltage: 2300 VAC
Max. Lamp Voltage: 300 VAC
Lamp Current: 18 mA
Demensions: 5.2"x2.2"x3.5" (132 mmx56 mmx89 mm)