How much turbidity can the Nitratax sc Sensor handle in the sample process?
Sommario
Nitratax sc and turbidity
Risposta
The Nitratax sc Sensor can handle turbidity up to 1000 NTU or FNU and still read reliably across all measurement pathlengths. The actual maximum turbidity value cannot be stated as it will
depend on the sludge structure.
Because the Nitratax sc sensor has two channels (measure and reference) and is working with wavelengths that are very close together, the sensor is able to see, allocate and compensate for turbidity up to 1000 NTU or FNU.