What is the relationship between ammonia (NH3) and other nitrogen containing compounds?
Sommario
Relationship between nitrogen containing compounds
Risposta
Nitrogen exists in the environment in many forms and changes forms as it moves through the nitrogen cycle. A variety of nitrogen methods can be used to test the different forms of nitrogen.
Provided below is a summary of the relationship between different nitrogen compounds and the nitrogen method recommended for analysis.
(TIN) = (NH3) + (NO3) + (NO2)
(TN) = Organic Nitrogen (all) + NH3 + NO3 + NO2
(TKN) = Organic Nitrogen (trinegative oxidative*) + NH3
*Trinegative oxidative: Will not account for azide, azine, azo, hydrazone, nitrile, nitro, nitroso, oxime, and semi-carbazone
Comparing TKN results
TKN = (approx) TN - (NO3 + NO2)
TKN = (approx) NH3 + Organic Nitrogen
Organic Nitrogen:
TKN - NH3 = Organic Nitrogen
TN-TIN = Organic Nitrogen
Acryonym Legend:
Total Inorganic Nitrogen (TIN)
Ammonia (NH3)
Ammonium (NH4)
Nitrate (NO3)
Nitrite (NO2)
Total Kjeldahl Nitrogen (TKN)
Total Nitrogen (TN)
Provided below is a summary of the relationship between different nitrogen compounds and the nitrogen method recommended for analysis.
(TIN) = (NH3) + (NO3) + (NO2)
(TN) = Organic Nitrogen (all) + NH3 + NO3 + NO2
(TKN) = Organic Nitrogen (trinegative oxidative*) + NH3
*Trinegative oxidative: Will not account for azide, azine, azo, hydrazone, nitrile, nitro, nitroso, oxime, and semi-carbazone
Comparing TKN results
TKN = (approx) TN - (NO3 + NO2)
TKN = (approx) NH3 + Organic Nitrogen
Organic Nitrogen:
TKN - NH3 = Organic Nitrogen
TN-TIN = Organic Nitrogen
Acryonym Legend:
Total Inorganic Nitrogen (TIN)
Ammonia (NH3)
Ammonium (NH4)
Nitrate (NO3)
Nitrite (NO2)
Total Kjeldahl Nitrogen (TKN)
Total Nitrogen (TN)